The TSi Power Corporation Home Page White Papers On Power Protection Strategies and Technologies Press Releases and News on TSi Power Corporation TSi Power & Surge Protection Products Overview Line Conditioners, Automatic Voltage Regulators, Low-Impedance and Ferroresonant Isolation Transformers Ac Automatic Voltage Regulators, Precision PWM, Tap-Switching and Ferroresonant Topologies Unipolar PWM Sine Wave Output Dc to Ac Inverters Automatic Transfer Switches For Enabling Redundant Power Sources e.g. 2 UPS Indoor Uninterruptible Power Systems Outdoor UPS With Extreme Wide Temperature Batteries for Harsh Environments Accessories for TSi Products, such as Rack-mounting Kits, Extra Battery Cabinets, etc. Pricing, RFQ's, RFP's and Getting Formal Quotations Frequently Asked Questions URL Links to some of TSi Power's Favorite Websites URL Link to the Free Adobe Acrobat PDF Reader Download Page

 * Best Viewed In Internet Explorer

 

ILc-Med Series
120vac, 1000,1400VA

Hospital-Grade UL 2601
Medical Isolation Transformer


       
.

ILc-1400Med4 1400VA, 120VAC, Medical Isolation Transformer For UL 2601 Applications

Click Here For Complete Specifications In 
A Printer Friendly Adobe Acrobat PDF Format

  ILc-1400Med4 Shown

TSi Power’s ILc-Med Series Medical Isolation Transformer Line Conditioners provide the best all-round hospital-grade 'low-leakage current' line conditioning and surge protection for sensitive medical equipment. The integral isolation transformer provides 100% isolation from the input AC line; the secondary neutral-ground bond eliminates all surge voltages between neutral and ground. Surge protection and noise filtering are superior to conventional surge protection and filtering devices.

Key ILc-Med Series Benefits
TSi Power’s Hospital- Grade Line Conditioner reduces the need for a high quality "isolated and dedicated" line, while eliminating interference from adjacent sources of disturbances. ILc-Med contributes to enhanced systems operation, greater reliability, improved repeatability, fewer reboots and system lock-ups and extended operating life. In addition, a reduction in hospital medical equipment down time and service costs can be expected.

How The ILc-Med Series Works
The integral isolation transformer provides complete isolation between the primary and secondary, and permits bonding its output neutral to ground, which completely eliminates all disturbances between neutral and ground, regardless of source. The series inductance of the isolation transformer in combination with a capacitive element and an MOV, provide superb noise filtering, as well as coordinated multi-stage surge protection in accordance with the principles of IEC-61312.3.

In addition, ILc-Med has an extremely low input AC leakage current (typical 50µA) rating which exceeds both domestic US (UL 2601), Canadian (cUL 2601) and international (IEC 601) medical standards for a leakage current limit of 300µA maximum for non-patient connected medical equipment & instruments.

Typical ILc-Med Applications
These are designed for medical systems requiring line conditioners having UL 2601 or Canadian cUL 2601 agency approval. Examples include: imaging systems, analyzers, patient care/ monitoring equipment, audio systems and other patient-critical applications.


Key Features Of The TSi Power ILc-Med Medical Isolation Transformer
Typical Leakage Current <50µA
UL and cUL 2601 Listed
Low-Impedance Isolation Transformer
Coordinated Surge Protection
Per IEC 61312.3
Filtering and MOV’s
Resettable Circuit Breaker

 

Eight "Green Dot" Medical Grade
Outlets/Receptacles
High Efficiency
100% Common-Mode Noise Elimination
True Isolation
Quiet Operation
5 Year Warranty

 

   SPECIFICATIONS ILc-1000Med4 ILc-1400Med4
    Electrical

 

   Capacity in VA (Watts) 1000 VA (1000 Watts) 1400 VA (1400 Watts)
     Transformer Type Hospital-Grade E-I Core Isolation Transformer
    Input
     Nominal Voltage 120 Volts AC, Single Phase
     Operating Voltage 100 - 130 Volts AC
     Nominal Frequency  60 Hz +/-5
     Leakage Current 50µA Typical (100µA Maximum)
     Over-Current Protection User resettable circuit breaker
     Circuit Breaker Rating 10 Amps 15 Amps
     AC Input Cord 15 Foot long hospital-grade AC power cord with NEMA 5-15P plug
    Output
     Nominal Voltage 120 Volts AC, Single Phase
     Power Efficiency 94% to 96%
     Total Harmonic Distortion < 1%
     Surge Protection A three stage surge protection system consisting of 
isolation transformer, capacitor and M.O.V. is included.
     Surge Test Conditions Per ANSI/IEEE C62.41-1991 test waveforms
     Surge Let-Through Voltages Single Pulse: L-N: 50V, L-G: 50V, N-G: 0.5V
Ring wave: L-N: 20V, L-G: 20V, N-G: 0.5V
Combination Wave: L-N: 250V, L-G: 250V, N-G: 0.5V
The measured rate of voltage rise/fall (dV/dt of the remnant waveform 
is less than 10V/
mS with input test waveform dV/dt of 6kV/mS.
     AC Receptacles 8 Hospital-Grade "Green Dot" NEMA 5-15R Receptacles
    Physical
     Dimensions 7.5"W x 5.25"H x 9.0"D
(190 x 133 x 229mm)
8.5"W x 6.5"H x 9.0"D
(216 x 165 x 229mm)
     Weight 29 lbs (13.2kg) 49 lbs (22.3kg)
    Safety
     Agency Approvals UL 2601 & cUL 2601 Listed.
UL Control Number 85EF
    Environmental
     Ambient Temperature 0 to +40°C, 90% Relative Humidity (non-condensing)
    Warranty
     Warranty 5 Year Limited Warranty, Parts & Labor
TSi's ongoing product improvement process makes specifications subject to change. Other companies product names herein are for identification purposes only, and may be trademarks of their respective companies. 

ILc-Med Series System Architecture  

                    ILc-Med Series, UL 2601 Medical Isolation Transformer System Architecture Drawing

ILc-Med Rear Views  

Rear View ILc-1000Med4, 1000VA, 120VAC, Medical Isolation Transformer For UL 2601 Applications Rear View ILc-1400Med4 1400VA, 120VAC, Medical Isolation Transformer For UL 2601 Applications
ILc-1000Med4 ILc-1400Med4

TSi Power also manufactures custom medical isolation transformers.
Please call if needed. 

Back To Top 

             

TSi Power Corporation ~ 1103 West Pierce Ave ~ Antigo, Wisconsin 54409 ~ Direct 715.623.0636 ~ 800.874.3160 (USA)
Fax 715.623.2426 ~ e-mail ~ sales@tsipower.com  ~  michaelg@carebase.com for comments/feedback on web site.
Privacy Statement    Copyright © 2001-2006 TSi Power Corporation

                                                        Hit Counter